Home Elpac by ICCNexergy 65 Watt Medical MWA065

65 Watt Medical MWA065

ICCNexergey MW065

Features:

  • High Efficiency: Level V
  • High Power Density 4.9W/in3
  • Lifetime Expectation >5 Years
  • Hold-up Time >25ms at Full Load
  • EISA, CEC Compliant
  • Medical Approval - EN60601-1 Class I
PEI can modify standard power supplies within 5 business days.

Modified options include:

  • Custom Connectors
  • Ferrite Cores
  • Custom Imprinting
  • Special Packaging/Labeling
  • Overmolding
  • Output Cord Lengths

Other factory options include:

  • Grounded Output
  • White Case

Model Overview

Elpac Part NumberOutput VoltageOutput CurrentPeak Current1Total Regulation2Typical Efficiency3
MWA065009A-11A 9.0V 8.0A 9.6A +/-5% 85%
MWA065012A-11A 12.0V 6.0A 7.2A +/-5% 88%
MWA065015A-11A 15.0V 4.8A 5.7A +/-5% 89%
MWA065024A-11A 24.0V 3.0A 3.6A +/-5% 91%
Input
Input Voltage 85 - 264VAC 100 - 240VAC Nominal
Input Frequency 47 – 63Hz
Input Current <1.5A rms
Inrush Current <37A at 230VAC cold start
Zero Load Power Consumption <0.5W
Earth Leakage Current (Typical)
<80µA @ 132VAC @ 60Hz
<100µA @ 264VAC @ 60Hz
Patient Leakage Current
<40µA @ 132VAC @ 60Hz
<50µA @ 264VAC @ 60Hz
Output
Output Voltage See Table
Total Regulation +/-5%
Minimum Load No minimum load required
Start-Up Delay <750ms
Hold-Up Time >25ms at any input voltage
Ripple & Noise <1% pk-pk *
Over Voltage Protection 110-135%
Over Temperature Protection Active - Recoverable; plus Passive - Non Recoverable
Over Current Protection 120 - 180%
Short Circuit Protection Shutdown, auto-restart (hiccup mode)
General
Efficiency Avg Efficiency 88.2% @ 115VAC; 89.7% @ 230VAC
MTBF min. 200,000 hours demonstrated
Size 5.02" (127.6mm) x 2.20" (55.9mm) x 1.37" (34.8mm)
Weight 1.3 lbs (.59 kg)
Power Density
4.9W/in3
Environmental
Operating Temperature 0 – 60°C (Full load to 40°C, derate linearly to 50% load at 60°C)
Storage Temperature -40°C to +85°C
Relative Humidity 5-95%, non-condensing
Cooling Natural Convection
Vibration All units production tested to 19.6m/s2
EMC & Safety
Emissions FCC class B, CISPR11 class B EN61000-3-2, -3
Immunity EN61000-4-2, -3, -4, -5, -6, -8, -11
Certified by:
cTUVus
UL 60601-1
CAN/CSA-22.2 No.601.1-M90*
CB per IEC60601-1
CE marked to LVD

Notes

  • 1 Maximum peak load (86W) lasting 500ms with a maximum 10% duty cycle.
  • 2 Includes initial setting, line regulation, load regulation, and thermal drift.
  • 3 Typical at 115VAC (including output cable).
  • ** Ripple and noise measured with 20MHz bandwidth; 10μF tantalum capacitor in parallel with a 0.1μF ceramic capacitor.

Input / Output Configuration

Input Configuration
Standard Input Cable Not Provided
Connection on Power Supply Body IEC 320 C14 Receptacle
Output Configuration
Standard Output Cable 6 ft.
Cord Size 2x16awg
Connector (PSU side) Switchcraft DIN-5, P/N 05GM5MX (male pins)
Mating Connector Switchcraft 57GB5FX or equivalent
Output Pin Assignments
DIN-5
Pin 1
Return
Pin 2
Return
Pin 3 +V1
Pin 4 Return
Pin 5 +V1

 
Welcome