Home Elpac by ICCNexergy 24 Watt Medical MW24

24 Watt Medical MW24

ICCNexergey MW24

Features:

  • High Efficiency
  • Fully Regulated DC Output
  • Lifetime Expectation >5 years
  • Hold-up Time >20ms at Full Load
  • Safety Approval - EN60601-1 Class I
PEI can modify standard power supplies within 5 business days.

Modified options include:

  • Custom Connectors
  • Ferrite Cores
  • Custom Imprinting
  • Special Packaging/Labeling
  • Overmolding
  • Output Cord Lengths

Other factory options include:

  • Grounded Output
  • White Case

Model Overview

Elpac Part NumberOutput VoltageOutput CurrentPeak Current1Total Regulation2Typical Efficiency3
MW2412-760-NC-BK 12.0V 2.0A 2.4A ±5% 80%
MW2415-760-NC-BK 15.0V 1.6A 1.9A ±5% 81%
MW2418-760-NC-BK 18.0V 1.3A
1.5A ±5% 81%
MW2424-760-NC-BK 24.0V 1.0A 1.2A ±5% 82%
Input
Input Voltage 85-264VAC; 100-240VAC Nominal
Input Frequency 47-63Hz
Input Current <0.5A rms
Inrush Current <37A at 230VAC cold start
Zero Load Power Consumption 0.85W
Touch Leakage Current
<100µA @ 132VAC @ 60Hz
<200µA @ 264VAC @ 60Hz
Output
Output Voltage See Model Overview Table
Total Regulation +/-5%
Minimum Load No minimum load required
Start-Up Delay <1s
Hold-Up Time >20ms
Ripple & Noise <1% pk-pk *
Over Voltage Protection 110-135%
Over Temperature Protection Active - Recoverable; plus Passive - Non Recoverable
Over Current Protection 120 - 180%
Short Circuit Protection Shutdown, auto-restart (hiccup mode)
General
Efficiency Avg Efficiency 81.5% @ 115VAC; 80.9% @ 230VAC
MTBF min. 200,000 hours demonstrated
Size 4.25" (108.0mm) x 2.5" (64.5mm) x 1.30" (33.0mm)
Weight 0.45 lbs (.20 kg)
Environmental
Operating Temperature 0 – 60°C (Full load to 40°C, derate linearly to 50% load at 60°C)
Storage Temperature -40°C to +85°C
Relative Humidity 5-95%, non-condensing
Cooling Natural Convection
Vibration All units production tested to 19.6m/s2
EMC & Safety
Emissions FCC class B, CISPR11 class B EN61000-3-2, -3
Immunity EN61000-4-2, -3, -4, -5, -6, -8, -11
Certified by:
cTUVus
UL 60601-1
CAN/CSA-22.2 No.601.1-M90*
CB per IEC60601-1
CE marked to LVD

Notes

  • 1 Maximum peak load (28.8W) lasting 500ms with a maximum 10% duty cycle.
  • 2 Includes initial setting, line regulation, load regulation, and thermal drift.
  • 3 Typical at 115VAC (including output cable).

Input / Output Configuration

Input Configuration
Standard Input Cable Not Provided
Connection on Power Supply Body IEC 320 C14 Receptacle
Output Configuration
Standard Output Cable 6 ft.
Cord Size 2x16awg zip
Connector (PSU side) Switchcraft 760 or equivalent
Mating Connector Switchcraft 712A or equivalent
Output Pin Assignments
Center +v1
Outside Return

 
Welcome
Please enter your Zip code/Country code to give you
up-to-the-minute pricing and availability in your area.

 I accept the terms and conditions of this website.