Home Elpac by ICCNexergy 12 Watt Medical MW12

12 Watt Medical MW12

MW12 Is Not Recommended For New Designs. Please Use MWA020 Class I As A Replacement.

ICCNexergey MW12

Features:

  • High Efficiency
  • Fully Regulated DC Output
  • Lifetime Expectation >5 years
  • Hold-up Time >20ms at Full Load
  • Safety Approval - EN60601-1 Class I
PEI can modify standard power supplies within 5 business days.

Modified options include:

  • Custom Connectors
  • Ferrite Cores
  • Custom Imprinting
  • Special Packaging/Labeling
  • Overmolding
  • Output Cord Length

Other factory options include:

  • Grounded Output
  • White Case

Model Overview

Elpac Part NumberOutput VoltageOutput CurrentPeak Current1Total Regulation2Typical Efficiency3
MW1212-760-NC-BK 12.0V 1.0A 1.2A ±5% 78%
MW1224-760-NC-BK 24.0V 0.5A 0.6A ±5% 78%
Input
Input Voltage 85 – 264VAC 100 – 240VAC Nominal
Input Frequency 47 – 63Hz
Input Current <0.5A rms
Inrush Current <37A at 230VAC cold start
Zero Load Power Consumption 0.97W
Touch Leakage Current
<100µA @ 132VAC @ 60Hz
<200µA @ 264VAC @ 60Hz
Output
Output Voltage See Table
Total Regulation +/-5%
Minimum Load No minimum load required
Start-Up Delay <1s
Hold-Up Time >20ms
Ripple & Noise <1% pk-pk ** *
Over Voltage Protection 110-135%
Over Temperature Protection Active - Recoverable; plus Passive - Non Recoverable
Over Current Protection 120 - 180%
Short Circuit Protection shutdown, auto-restart (hiccup mode)
General
Efficiency Avg Efficiency 78.1% @ 115VAC; 77.3% @ 230VAC
MTBF min. 200,000 hours demonstrated
Size 3.35" (85.0mm) x 1.97" (50.0mm) x 1.30" (33.0mm)
Weight 0.45 lbs (.20 kg)
Environmental
Operating Temperature 0 – 60°C (Full load to 40°C, derate linearly to 50% load at 60°C)
Storage Temperature -40°C to +85°C
Relative Humidity 5-95%, non-condensing
Cooling Natural Convection
Vibration All units production tested to 19.6m/s2
EMC & Safety
Emissions FCC class B, CISPR11 class B EN61000-3-2, -3
Immunity EN61000-4-2, -3, -4, -5, -6, -8, -11
Certified by:
cTUVus
UL 60601-1
CAN/CSA-22.2 No.601.1-M90*
CB per IEC60601-1
CE marked to LVD

Notes

  • 1 Maximum peak load (14.4W) lasting 500ms with a maximum 10% duty cycle.
  • 2 Includes initial setting, line regulation, load regulation, and thermal drift.
  • 3 Typical at 115VAC (including output cable).
  • * Ripple and noise measured with 20MHz bandwidth; 10μF tantalum capacitor in parallel with a 0.1μF ceramic capacitor.

Input / Output Configuration

Input Configuration
Standard Input Cable Not Provided
Connection on Power Supply Body IEC 320 C14 Receptacle
Output Configuration
Standard Output Cable 6 ft.
Cord Size 2x16awg zip
Connector (PSU side) Switchcraft 760 or equivalent
Mating Connector Switchcraft 712A or equivalent
Output Pin Assignments
Center +v1
Outside Return

 