Home Elpac by ICCNexergy

Elpac by ICCNexergy 100W AC-DC Medical Open Frame Power Supply – MVA100 Series

100W OPEN FRAME POWER RATED TO EN60601-1 CLASS I 3RD EDITION

The Elpac by ICCNexergy 100W AC-DC Medical Open Frame Power Supply – MVA100 Series offers power output of 100 watts (up to 120 watts with forced air cooling) with standard output voltages of 12V, 15V, 18V and 24V. This 100W internal power supply accepts universal input voltage of 85 – 264 VAC. The MVA100 Series meets Medical Approval EN60601-1 Class I 3rd Edition. EN60601-1 Class II is available from the factory. For full details on the Elpac by ICCNexergy 100W AC-DC Medical Open Frame Power Supply – MVA100 Series, please see the product specifications below.

Click to download PDF

FEATURES

ELPAC BY ICCNEXERGY 100W AC-DC MEDICAL OPEN FRAME POWER SUPPLY – MVA100 SERIES FEATURES

  • High Efficiency: Level V

  • Up to 120W with Forced Air

  • High Power Density 8.7W/in3

  • Hold-up Time > 24ms at full load

  • EISA & CEC Compliant

  • Output Floating

  • Medical Approval - EN60601-1 Class I 3rd Edition

  • Class II available, consult factory


PART NUMBER

ELPAC BY ICCNEXERGY 100W AC-DC MEDICAL OPEN FRAME POWER SUPPLY – MVA100 SERIES PART NUMBER

Elpac
Part Number

Output
Voltage

Output
Current1

Peak
Current

Total
Regulation2

Typical Efficiency3

MVA100012A

12.0V

8.3A

10.0A

±5%

90%

MVA100015A

15.0V

6.7A

8.0A

±5%

90%

MVA100018A

18.0V

5.6A

6.7A

±5%

91%

MVA100024A

24.0V

4.2A

5.0A

±5%

92%


Notes:
1 With convection cooling. Peak load (120W) lasting 500ms with a maximum 10% duty cycle
2 Includes initial setting, line regulation, load regulation, and thermal drift
3 Typical at 115VAC and full load

INPUT

ELPAC BY ICCNEXERGY 100W AC-DC MEDICAL OPEN FRAME POWER SUPPLY – MVA100 SERIES INPUT

Input

Input Voltage

85 - 264VAC 100 - 240VAC Nominal

Input Frequency

47 - 63Hz

Input Current

< 2A rms

Inrush Current

< 37A at 230VAC cold start

Zero Load Power Consumption

<0.5W

Patient Leakage Current

<50µA @ 132VAC @ 60Hz
<80µA @ 264VAC @ 60Hz

Touch Leakage Current

<125µA @ 132VAC @ 60Hz
<175µA @ 264VAC @ 60Hz

OUTPUT

ELPAC BY ICCNEXERGY 100W AC-DC MEDICAL OPEN FRAME POWER SUPPLY – MVA100 SERIES OUTPUT

Output

Output Voltage

See Part Number Table

Total Regulation

+/-5% (each output)

Minimum Load

No minimum load required

Start-Up Delay

<750ms

Hold-Up Time

>24ms at any input voltage

Ripple & Noise

< 1% pk-pk *

Over Voltage Protection

110-135%

Over Temperature Protection

Active - Recoverable; plus Passive - Non Recoverable

Over Current Protection

120 - 180%

Short Circuit Protection

shutdown, auto-restart (hiccup mode)


Note:

* Ripple and noise measured with 20MHz bandwidth; 10μF tantalum capacitor in parallel with a 0.1μF ceramic capacitor.

GENERAL

ELPAC BY ICCNEXERGY 100W AC-DC MEDICAL OPEN FRAME POWER SUPPLY – MVA100 SERIES GENERAL SPECS

General

Efficiency

Avg Efficiency 88.6% @ 115VAC; 89.5% @ 230VAC

MTBF

min. 200,000 hours demonstrated

Size

4.50" x 2.5" x 1.22" | 114mm x 63.5mm x 30.9mm

Weight

0.45 lbs (.20 kg)

Power Density

8.7W/in3

ENVIRONMENTAL

ELPAC BY ICCNEXERGY 100W AC-DC MEDICAL OPEN FRAME POWER SUPPLY – MVA100 SERIES ENVIRONMENTAL

Environmental

Operating Temperature

0 – 70°C (Full load to 50°C, derate linearly to 50% load at 70°C)

Storage Temperature

-40°C to +85°C

Relative Humidity

5-95%, non-condensing

Cooling

Natural Convection (100W) or Forced Air (120W)

Vibration

All units production tested to 19.6m/s2

EMC & SAFETY

ELPAC BY ICCNEXERGY 100W AC-DC MEDICAL OPEN FRAME POWER SUPPLY – MVA100 SERIES EMC & SAFETY

EMC & Safety

Emissions

FCC class B, CISPR11 class B EN61000-3-2, -3

Immunity

EN61000-4-2, -3, -4, -5, -6, -8, -11



Certified by TUV to the following:



cTUVus
UL 60601-1
CAN/CSA-22.2 No.601.1-M90
CB per IEC60601-1 2nd and 3rd Edition
CE marked to LVD

 
Welcome