Home Elpac by ICCNexergy

Elpac by ICCNexergy 40W AC-DC Medical Open Frame Power Supply – MS40 Series

40W OPEN FRAME POWER RATED TO EN60601-1 CLASS I

The Elpac by ICCNexergy 40W AC-DC Medical Open Frame Power Supply – MS40 Series offers power output of 40 watts with standard output voltages of 9V, 12V, 15V, 18V and 24V. This 40W internal power supply accepts universal input voltage of 85 – 264 VAC. The MS40 series meets Medical Approval EN60601-1 Class I. For full details on the Elpac by ICCNexergy 40W AC-DC Medical Open Frame Power Supply – MS40 Series, please see the product specifications below.

Click to download PDF

FEATURES

ELPAC BY ICCNEXERGY 40W AC-DC MEDICAL OPEN FRAME POWER SUPPLY – MS40 SERIES FEATURES

  • High Efficiency

  • Wide Range AC Input

  • Fully Regulated DC Output

  • Lifetime Expectation > 5 years

  • Hold-up Time > 16ms at full load

  • Output Floating

  • Safety Approval - EN60601-1 Class I

  • RoHs Compliant


PART NUMBER

ELPAC BY ICCNEXERGY 40W AC-DC MEDICAL OPEN FRAME POWER SUPPLY – MS40 SERIES PART NUMBERS

Elpac Part Number

Output Voltage

Output Current1

Total Regulation2

Typical Efficiency3

MS4009F

9.0V

4.40A

±5%

84%

MS4012F

12.0V

3.30A

±5%

86%

MS4015F

15.0V

2.70A

±5%

87%

MS4018F

18.0V

2.20A

±5%

86%

MS4024F

24.0V

1.67A

±5%

88%


Notes:
1 Maximum peak load (48W) lasting 500ms with a maximum 10% duty cycle
2 Includes initial setting, line regulation, load regulation, and thermal drift
3 Typical at 115VAC


INPUT

ELPAC BY ICCNEXERGY 40W AC-DC MEDICAL OPEN FRAME POWER SUPPLY – MS40 SERIES INPUT

Input

Input Voltage

85 - 264VAC 100 - 240VAC Nominal

Input Frequency

47 - 63Hz

Input Current

< 1.0A rms

Inrush Current

< 37A at 230VAC cold start

Zero Load Power Consumption

< 0.5W

Earth Leakage Current (Typical)

< 50µA @ 132VAC @ 60Hz
< 75µA @ 264VAC @ 60Hz

OUTPUT

ELPAC BY ICCNEXERGY 40W AC-DC MEDICAL OPEN FRAME POWER SUPPLY – MS40 SERIES OUTPUT

Output

Output Voltage

See Part Number Table

Total Regulation

+/-5%

Minimum Load

No minimum load required

Start-Up Delay

< 150ms

Hold-Up Time

> 16ms

Ripple & Noise

< 1% pk-pk *

Over Voltage Protection

130 - 135%

Over Temperature Protection

Active - Recoverable; plus Passive - Non Recoverable

Over Current Protection

120 - 180%

Short Circuit Protection

shutdown, auto-restart (hiccup mode)


Note:
* Ripple and noise measured with 20MHz bandwidth; 10μF tantalum capacitor in parallel with a 0.1μF ceramic capacitor.

GENERAL

ELPAC BY ICCNEXERGY 40W AC-DC MEDICAL OPEN FRAME POWER SUPPLY – MS40 SERIES GENERAL SPECS

General

Efficiency

Avg Efficiency 87.3% @ 115VAC; 86.9% @ 230VAC

MTBF

min. 200,000 hours demonstrated

Size

3.0" x 5.0" x 1.1" | 76.2mm x 127.0mm x 27.5mm

Weight

0.40 lbs (0.18 kg)

Power Density

2.9W/in3

ENVIRONMENTAL

ELPAC BY ICCNEXERGY 40W AC-DC MEDICAL OPEN FRAME POWER SUPPLY – MS40 SERIES ENVIRONMENTAL

Environmental

Operating Temperature

0 – 50°C (Full load to 50°C, derate linearly to 50% load at 70°C)

Storage Temperature

-40°C to +85°C

Relative Humidity

5 - 95%, non-condensing

Cooling

Natural Convection

Vibration

All units production tested to 19.6m/s2

EMC & SAFETY

ELPAC BY ICCNEXERGY 40W AC-DC MEDICAL OPEN FRAME POWER SUPPLY – MS40 SERIES EMC & SAFETY

EMC & Safety

Emissions


Immunity




Certified by UL to the following:



cRUus
UL 60601-1
CAN/CSA-22.2 No.601.1-M90
CB per IEC60601-1
CE marked to LVD

 