Home Elpac by ICCNexergy 20 Watt (Class I) MWA020

20 Watt (Class I) MWA020

ICCNexergey MW20

Features:

  • High efficiency
  • Fully regulated DC output
  • Lifetime expectation >5 years
  • Hold-up time >20ms at full load
  • Safety approval - EN60601-1 Class I
PEI can modify standard power supplies within 5 business days.

Modified options include:

  • Custom connectors
  • Ferrite cores
  • Custom imprinting
  • Special packaging/labeling
  • Overmolding
  • Output cord length

Other factory options include:

  • Grounded output
  • White case

Model Overview

Elpac Part NumberOutput VoltageOutput CurrentPeak Current1Total Regulation2Typical Efficiency3
MWA020005A-10A 5.0V 4.0A 4.2A ±5% 80%
MWA020009A-10A
9.0V
2.22A 2.3A ±5% 83%
MWA020012A-10A 12.0V 1.67A
1.75A ±5% 84%
MWA020024A-10A 24.0V 0.84A 0.87A ±5% 86%
Input
Input Voltage 85-264VAC; 100-240VAC Nominal
Input Frequency 47-63Hz
Input Current <0.5A rms
Inrush Current <50A @ 230VAC cold start
Zero Load Power Consumption <0.3W
Earth Leakage Current (Typical)
<100µA @ 132VAC @ 60Hz (typical)
<150µA @ 264VAC @ 60Hz (typical)
Output
Output Voltage See Model Overview Table
Total Regulation +/-5%
Minimum Load No minimum load required
Start-Up Delay <1.5s
Hold-Up Time >18ms at any input voltage
Ripple & Noise <1% pk-pk**
Over Voltage Protection 112-135%
Over Temperature Protection Active - Recoverable; plus Passive - Non Recoverable
Over Current Protection 108% - 180%
Short Circuit Protection Shutdown, auto-restart (hiccup mode)
General
Efficiency Avg Efficiency 83.4% at 115VAC; 82.8% @ 230VAC
MTBF min. 200,000 hours demonstrated
Size 3.45" x 2.04" x 1.30" | 87.7mm x 5.18mm x 33.0mm
Weight 5V 0.4lbs (0.2Kg); 24V 0.3 lbs (0.16Kg)
Power Density
3.0 W/in³
Environmental
Operating Temperature 0 – 60°C (Full load to 40°C, derate linearly to 50% load at 60°C)
Storage Temperature -40°C to +85°C
Relative Humidity 5-95%, non-condensing
Cooling Natural Convection
Vibration All units production tested to 19.6m/s2
EMC & Safety
Emissions FCC class B, CISPR11 class B EN61000-3-2, -3
Immunity EN61000-4-2, -3, -4, -5, -6, -8, -11
Certified by TUV to the following:
cTUVus
UL 60950-1
CAN/CSA-22.2 No.601
CB per IEC60601-1
CE marked to LVD Directive

Notes

  • 1 Maximum peak load (21W) lasting 500ms with a maximum 10% duty cycle.
  • 2 Includes initial setting, line regulation, load regulation, and thermal drift.
  • 3 Typical at 115VAC (including output cable).

Input / Output Configuration

Input Configuration
Standard Input Cable Not Provided
Connection on Power Supply Body IEC 320 C14 Receptacle
Output Configuration
Standard Output Cable 6 ft.
Cord Size 16AWG zip (5V, 9V)
18AWG zip (12V)
20AWG zip (24V)
Connector (PSU side) Switchcraft 760 or equivalent
Mating Connector Switchcraft 712A or equivalent
Output Pin Assignments
Center +v1
Outside Return

 
Welcome
Please enter your Zip code/Country code to give you
up-to-the-minute pricing and availability in your area.

 I accept the terms and conditions of this website.