Home Elpac by ICCNexergy 40 Watt Medical Open Frame MTA040

40 Watt Medical Open Frame MTA040

Features:

  • High Efficiency: Level V
  • Fully Regulated DC Output
  • Lifetime Expectation >5 years
  • Hold-up Time >14ms at Full Load
  • EISA & CEC Compliant
  • Floating Output
  • Safety Approval - EN60601-1 Class I
PEI can modify standard power supplies within 5 business days.

Model Overview

Elpac Part NumberOutput VoltageOutput Current1Total Regulation2Typical Efficiency3
MTA040009A 9.0V 4.40A ±5% 85%
MTA040012A 12.0V 3.30A ±5% 87%
MTA040015A 15.0V
2.70A
±5% 88%
MTA040024A 24.0V 1.70A ±5% 87%
Input
Input Voltage 85 – 264VAC 100 – 240VAC Nominal
Input Frequency 47 – 63Hz
Input Current <1.0A rms
Inrush Current <37A at 230VAC cold start
Zero Load Power Consumption <0.3W
Earth Leakage Current (Typical)
<60µA @ 132VAC @ 60Hz
<80µA @ 264VAC @ 60Hz
Patient Leakage Current
<30µA @ 132VAC @ 60Hz
<50µA @ 264VAC @ 60Hz
Output
Output Voltage See Table
Total Regulation +/-5%
Minimum Load No minimum load required
Start-Up Delay <250ms
Hold-Up Time >14ms
Ripple & Noise <1% pk-pk ** *
Over Voltage Protection 110-135%
Over Temperature Protection Active - Recoverable; plus Passive - Non Recoverable
Over Current Protection 120 - 180%
Short Circuit Protection shutdown, auto-restart (hiccup mode)
General
Efficiency Avg Efficiency 87.1% @ 115VAC; 87.1% @ 230VAC
MTBF min. 200,000 hours demonstrated
Size 4.00" x 2.00" x 0.99" | 101.6mm x 50.81mm x 25.2mm
Weight 0.29 lbs (0.13 kg)
Power Density
5.76W/in3
Environmental
Operating Temperature 0 – 50°C (Full load to 50°C, derate linearly to 50% load at 70°C)
Storage Temperature -40°C to +85°C
Relative Humidity 5-95%, non-condensing
Cooling Natural Convection
Vibration All units production tested to 19.6m/s2
EMC & Safety
Emissions FCC class B, CISPR11 class B EN61000-3-2, -3
Immunity EN61000-4-2, -3, -4, -5, -6, -8, -11
Certified by TUV to the following:
cTUVus
UL 60601-1
CAN/CSA-22.2 No.601.1-M90*
CB per IEC60601-1
CE marked to LVD

Notes

  • 1 Maximum peak load (48W) lasting 500ms with a maximum 10% duty cycle.
  • 2 Includes initial setting, line regulation, load regulation, and thermal drift.
  • 3 Typical at 115VAC
  • * Ripple and noise measured with 20MHz bandwidth; 10μF tantalum capacitor in parallel with a 0.1μF ceramic capacitor.

Input / Output Configuration

Input Configuration
Connection on Power Supply Body   
AMP p/n 640445-3 (or equivalent)
Mating Connector   
AMP p/n 640250-3 (or equivalent)
Output Configuration
Connector (PSU Side)   
AMP p/n 640445-6 (or equivalent)
Mating Connector AMP p/n 640250-6 (or equivalent)
Input Pin Assignments (P1)
Pin 1 AC Line
Pin 2 <not assembled>
Pin 3
AC Neutral
Output Pin Assignments (P2)
Pin 1 +V1
Pin 2 +V1
Pin 3
Returns
Pin 4
Return
Pin 5
No Connection
Pin 6
No Connection

 